Specifications

ZVx DISPlay Subsystem
1043.0009.50 3.47 E-15
DISPlay[:WINDow[1...4]]:DIAGram:SEGMented:R
This command defines the segment limits for polar diagrams. A maximum of 3 segments can be
defined. The separation line between two segments is common, i.e. there are neither gaps nor
overlaps. The numeric values refer to the unit used in the diagram and are sorted in descending
order.
Syntax:
DISPlay[:WINDow[1...4]]:DIAGram:SEGMented:R <numeric_value> ...
Example:
"DISP:DIAG:SEGM:R 20,-30,-70,-120"
Features:
*RST value:
SCPI:
-
device-specific
DISPlay[:WINDow[1...4]]:DIAGram:SEGMented:Y
This command defines the segment limits for Cartesian diagrams. A maximum of 3 segments can be
defined. The separation line between two segments is common, i.e. there are neither gaps nor
overlaps. The numeric values refer to the unit used in the diagram and are sorted in descending
order.
Syntax:
DISPlay[:WINDow[1...4]]:DIAGram:SEGMented:Y <numeric_value> ...
Example:
"DISP:DIAG:SEGM:Y 20,-30,-70,-120"
Features:
*RST value:
SCPI:
-
device-specific
DISPlay[:WINDow[1...4]]:TRACe2:X:OFFSet
This command sets the stimulus offset of a memory trace. It is therefore only available for TRAC2.
Syntax:
DISPlay[:WINDow[1...4]]:TRACe2:X:OFFSet <numeric_value>
Example:
"DISP:TRAC2:X:OFFs 10MHZ"
Features:
*RST value:
SCPI:
0 Hz
device-specific
DISPlay[:WINDow[1 to 4]]:TRACe[1|2]:X:SPACing
This command toggles between linear and logarithmic display of the X-axis.
Syntax: DISPlay[:WINDow[1 to 4]]:TRACe[1|2]:X:SPACing LINear | LOGarithmic | dB
Example: "DISP:TRAC:X:SPAC LOG"
Features: *RST value: LINear
SCPI: conforming
DISPlay[:WINDow[1 to 4]]:TRACe[1|2]:Y[:SCALe]:AUTO
This command performs a single rescaling of the y-axis or the radial axis, respectively.
Syntax: DISPlay[:WINDow[1 to 4]:TRACe[1|2]:Y[:SCALe]:AUTO ONCE
Example: "DISP:TRAC:Y:AUTO ONCE"
Features: *RST value:
SCPI: conforming
This command is an "event", which is why it is not assigned an *RST value and has no query.