Specifications
CALCulate Subsystem ZVx
1043.0009.50 3.34 E-15
CALCulate[1...4]:MARKer[1...8]:FUNCtion:QFACtor
This command defines the quality factor for the marker search function.
Syntax:
CALCulate[1...4]:MARKer[1...8]:FUNCtion:QFACtor
Example:
"CALC:MARK:FUNC:QFAC 100"
Features:
*RST value:
SCPI:
–
device-specific
CALCulate[1...4]:MARKer[1...8]:FUNCtion:SFACtor
This command defines the form factor for the marker search function.
Syntax:
CALCulate[1...4]:MARKer[1...8]:FUNCtion:SFACtor <numeric_value>,
<numeric_value>
Example:
"CALC:MARK:FUNC:SFAC 60dB,3dB"
Features:
*RST value:
SCPI:
–
device-specific
CALCulate[1...4]:MARKer[1...8]:FUNCtion:TARGet
This command defines the target value for the fixed-value search mode.
Syntax:
CALCulate[1...4]:MARKer[1...8]:FUNCtion:TARGet <numeric_value>
Example:
"CALC:MARK:FUNC:TARG 1.75"
Features:
*RST value:
SCPI:
–
device-specific
CALCulate[1...4]:MARKer[1...8]:FUNCtion:RESULT?
This command queries the result of the marker search function.
Syntax:
CALCulate[1...4]:MARKer[1...8]:FUNCtion:RESULT?
Example:
"CALC:MARK:FUNC:RES?"
Features:
*RST value:
SCPI:
-
device-specific
This command is a query,, which is why it is not assigned an *RST value.
CALCulate[1...4]:MARKer[1...8]:FUNCtion:EDELay
This command switches on or off the display of the electrical or the mechanical length or the phase
delay. The suffix of MARKer has no meaning. The parameter information means the following:
ELENgth = electrical length
DISTance = mechanical length
TIME = phase delay
OFF = display switched off
Syntax:
CALCulate[1...4]:MARKer[1...8]:FUNCtion:EDELay TIME | DISTance |
ELENgth | OFF
Example:
"CALC:MARK:FUNC:EDEL TIME"
Features:
*RST value:
SCPI:
OFF
device-specific