Specifications

CALCulate Subsystem ZVx
1043.0009.50 3.26 E-15
CALCulate[1..4]:LIMit[1...8]:CONTrol:SHIFt
This command shifts a limit line along the x-axis direction by the value specified.
Syntax: CALCulate<1|2>:LIMit<1...8>:CONTrol:SHIFt <numeric_value>
Example: "CALC:LIM2:CONTrol:SHIFT 50KHZ"
Features: *RST value: --
SCPI: device-specific
The command is an "event",, which is why it is not assigned an *RST value.
CALCulate[1...4]:LIMit[1...8]:CENTer
This command defines the coordinates of the center of the tolerance circle. .
Syntax:
CALCulate[1...4]:LIMit[1...8]:CENTer <numeric_value>,<numeric_value>
Example:
"CALC:LIM:CENT 0,0"
Features:
*RST value:
SCPI:
-
device-specific
CALCulate[1...4]:LIMit[1...8]:CENTer:SHIFt
This command shifts the tolerance circle center.
Syntax:
CALCulate[1...4]:LIMit[1...8]:CENTer:SHIFt <numeric_value>,<numeric_value>
Example:
"CALC:LIM:CENT:SHIFt 0.5,0.5"
Features:
*RST value:
SCPI:
-
device-specific
CALCulate[1...4]:LIMit[1...8]:UPPer[:DATA]
This command defines the values for the upper limit lines.
Syntax:
CALCulate[1...4]:LIMit[1...8]:UPPer[:DATA <numeric_value>,
<numeric_value>..
Example:
"CALC:LIM:UPP -10,0,0,-10"
Features:
*RST value:
SCPI:
-
conforming
The number of values for the CONTrol-axis and the corresponding UPPer limit line must be identical.
If the measured values exceed the UPPer limit line, the limit test reports an error.
CALCulate[1...4]:LIMit[1...8]:UPPer:SHIFt
This command shifts the tolerance band along the y-axis direction.
Syntax:
CALCulate[1...4]:LIMit[1...8]:UPPer:SHIFt <numeric_value>
Example:
"CALC:LIM:UPPer:SHIFt 3dB"
Features:
*RST value:
SCPI:
-
device-specific
The LOWer limit is shifted together with the UPPer limit by the same distance.