Specifications

Switchover to Remote Control ZVx
1043.0009.50 3.2 E-15
3.3 Switchover to Remote Control
On power-on, the instrument is always in the manual operating state ("LOCAL" state) and can be
operated via the front panel.
It is switched to remote control ("REMOTE" state)
IEC/IEEE bus as soon as it receives an addressed command from a controller.
RS-232 as soon as it receives the command "@REM" from a controller.
RSIB as soon as it receives an addressed command from a controller.
During remote control, operation via the front panel is disabled. The instrument remains in the remote
state until it is reset to the manual state via the front panel or via remote control interfaces. Switching
from manual operation to remote control and vice versa does not affect the remaining instrument
settings.
3.3.1
3.3.1 Remote Control via IEC Bus
3.3.1.1 Setting the Device Address
In order to operate the instrument via the IEC/IEEE bus Ι, it must be addressed using the set IEC/IEEE
bus address. The IEC/IEEE bus address of the instrument is factory-set to 20. It can be changed
manually in the SETUP - GENERAL SETUP menu or via IEC bus. Addresses 0 to 30 are permissible.
Manually: ½ Call SETUP - GENERAL SETUP menu
½ Enter desired address in table GPIB ADDRESS
½ Terminate input using one of the unit keys (=ENTER).
Via IEC bus:
CALL IBFIND("DEV1", analyzer%) Open port to the instrument
CALL IBPAD(analyzer%, 20) Inform controller about old address
CALL IBWRT(analyzer%, "SYST:COMM:GPIB:ADDR 18")Set instrument to new address
CALL IBPAD(analyzer%, 18) Inform controller about new address
3.3.1.2 Indications during Remote Control
Remote control mode is indicated by the LED "REMOTE" "on the instruments front panel. In this mode
the softkeys on the display are not shown.