Specifications

Performance Test Record (ZVK) ZVx
1127.8500.60 5.70 E-2
5.4 Performance Test Record (ZVK)
The indicated uncertainties refer to the proposed test setup / test procedure.
The expanded measurement uncertainty amounts to k = 2 (confidence level of 95%, Gauss distribution).
Additional measurement uncertainties, which are within the users responsibility, have not been taken
into account (e.g. cable attenuation in harmonics measurements).
We recommend performing a measurement uncertainty analysis for the test setup in question so that
the values specified in the data sheet can be reliably checked.
Table 5-2: Performance Test Record Generator Specifications
Item Characteristic Measure-
ment acc.
to section
Specification
min. value/Hz
Measured
value/Hz
Specification
max.
value/Hz
Uncer-
tainty / Hz
1 Frequency accuracy
0.010 GHz
0.100 GHz
0.5 GHz
1.5 GHz
1.998 GHz
3 GHz
5 GHz
8 GHz
15 GHz
20 GHz
25 GHz
30 GHz
36 GHz
40 GHz
5.2.1.1
-40
-400
-2000
-6000
-7992
-12000
-20000
-32000
-60000
-80000
-100000
-120000
-144000
-160000
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
____________
+40
+400
+2000
+6000
+7992
+12000
+20000
+32000
+60000
+80000
+100000
+120000
+144000
+160000
3
26
129
386
514
772
1286
2058
3858
5143
6429
7715
9258
10287