Specifications
ZVx
1127.8500.60 5.17 E-2
5.2.3.5 Dynamic Range
Test equipment: ZVM: 2 PC 3.5 SHORT (e.g. SHORT male and SHORT female with THRU male
from Calibration Kit ZV-Z32);
ZVK: 2 PC 2.92 SHORT (e.g. SHORT male and SHORT female with THRU male
from Calibration Kit ZV-Z34)
Test setup: Connect short-circuits to PORT1 and PORT2.
Settings on network analyzer:
PRESET
SWEEP LIN SWEEP
FORMAT MAGNITUDE
SOURCE POWER Maximum value acc. Data Sheet
Note: For the maximum value, check if the instru-
ment is fitted with a generator step attenuator which
may be different for the two channels.
ZVM 5 dBm to -2 dBm, ZVK 0 dBm to -9 dBm
IF BANDWIDTH 10 Hz or 10 kHz
AVERAGE POINT AVG 5
SMOOTHING 1%
MEAS RATIO b1/b2, DRIVE PORT 2 (dyn. range Port 1)
RATIO b2/b1, DRIVE PORT 1 (dyn. range Port 2)
MARKER Marker frequency = measurement frequency
Measurement: ½ Read marker value on network analyzer.