Specifications
Test Procedure (ZVM & ZVK) ZVx
1127.8500.60 5.8 E-2
5.2.1.7 Level Linearity
Test equipment: FSEK30, test cable, BNC cable for reference frequency
Test setup 50Ω: ½ Connect FSEK30 to PORT1 (PORT2) of the network analyzer. Synchronize
FSEK30 with network analyzer.
½ Important: IF BW on FSEK30 < 1 kHz (digital bandwidths)
Settings on network analyzer:
PRESET
SWEEP SINGLE POINT
CENTER Measurement frequency
SOURCE POWER Min. to max. value -20 dBm to -2/2/5 dBm, ZVM
Min. to max. value -20 dBm to -9/-5/-3/0dBm, ZVK
SWEEP TIME 255 s
MEAS INPUT a1
INPUT a2
Measurement:
½ Set measurement frequencies and levels according to test record, read level on
spectrum analyzer.