Specifications

ZVx Test Procedure (ZVM & ZVK)
1127.8500.60 5.7 E-2
5.2.1.6 Level Accuracy
Test equipment: Power Meter NRVD with sensor NRV-Z52 (ZVM), NRV-Z55 (ZVK),
Female-female adapter from calibration kit ZV-Z32 (ZVM) or ZV-Z34 (ZVK)
Test setup: ½ Connect power sensor to PORT1, PORT2 of the network analyzer.
Settings on network analyzer:
PRESET
(corresponds to a set generator level of 10 dBm)
SWEEP SINGLE POINT
CENTER Measurement frequency
SWEEP TIME 255 s
MEAS INPUT a1 (PORT1), INPUT a2 (PORT2)
Measurement: Set measurement frequencies according to test record, read level on power meter.
Add the correction value associated with the measurement frequency to the meas-
ured value (typ. attenuation of adapter).
Correction
value:
Frequency
in GHz
0.010
0.100
0.150
0.500
1.000
1.500
2.000
3.000
4.000
5.000
6.000
7.000
8.000
9.000
10.00
11.00
12.00
13.00
14.00
15.00
16.00
17.00
18.00
19.00
20.00
22.00
24.00
26.00
28.00
30.00
32.00
34.00
36.00
38.00
40.00
Correction value
ZVM
in dB
0.003
0.009
0.011
0.021
0.029
0.036
0.041
0.050
0.058
0.065
0.071
0.077
0.082
0.087
0.092
0.097
0.101
0.105
0.109
0.113
0.117
0.120
0.124
0.127
0.130
Correction value
ZVK
in dB
0.002
0.008
0.009
0.017
0.024
0.029
0.034
0.041
0.047
0.053
0.058
0.063
0.067
0.071
0.075
0.079
0.082
0.086
0.089
0.092
0.095
0.098
0.101
0.103
0.106
0.111
0.116
0.121
0.125
0.130
0.134
0.138
0.142
0.146
0.150