Specifications

ZVx Test Procedure (ZVM & ZVK)
1127.8500.60 5.5 E-2
5.2.1.4 Phase Noise
Test equipment:
Modulation meter FMB with option FMA-B8, FSEK30, BNC connecting cable
Test setup: ½ Connect modulation meter (mode DEMOD PM PHASENOISE 10 kHz) to PORT1
of the network analyzer.
Settings on network analyzer:
PRESET
SWEEP SINGLE POINT
CENTER Measurement frequency
SOURCE POWER 2 dBm (ZVM), -9 dBm (ZVK)
SWEEP TIME 255 s
MEAS INPUT a1
Measurement:
Set measurement frequencies according to test record, read phase noise on
modulation meter.