Specifications
Test Sequence (R&S ZVC, R&S ZVCE) ZVx
1043.0009.50 5.52 E-7
Reference
measurement:
One-port calibration including THRU over the complete frequency range (for
measurement at b1 and b2 an extra calibration each).
Measurement:
Set marker frequencies according to test report, measure b1 and b2.
Record return loss values.
Return loss: > 8 dB
Alternative Measurement (firmware version 1.50 and higher):
Test equipment,
test set 50 :
Calibration kit R&S ZV-Z21, pair of test cables R&S ZV-Z11
Test setup: Connect test PORT2 to Input b1 or PORT1 to Input b2 via test cable.
PRESET
SWEEP LOG SWEEP
MARKER Test frequency
MEAS S22 for measurement Input b1
S11 for measurement Input b2
SERVICE FUNCTION 2.13.1.2.2 for measurement Input b1 (reset: 2.13.0)
2.13.1.4.2 for measurement Input b2 (reset: 2.13.0)
Calibration:
Perform one-port calibration at the end of the test cable (towards Input b1 or Input b2).
Measurement:
Set test frequencies according to test report, measure b1 and b2.
Take return loss values.
Return loss 20 kHz
1)
to 8 GHz: > 8 dB
1)
Measurement below 20 kHz not required because the matching value is uniquely determined by the design at low frequencies.
A possible fault is recognized when the level accuracy is measured.