Specifications

Test Sequence (R&S ZVC, R&S ZVCE) ZVx
1043.0009.50 5.50 E-7
60 dB to 70 dB (ZVCE) <1 dB <6°
60 dB to 75 dB (ZVC) <1 dB <6°
4
GHz to 8 GHz +10 dB to +3 dB <1 dB <6°
+ 3 dB to 45 dB <0.2 dB <2°
45 dB to 55 dB (ZVCE) <1 dB <6°
45 dB to 60 dB (ZVC) <1 dB <6°
Model 51, 52, 61, 62: Deviation from
reference value (–10 dBm): Permissible deviation:
300 kHz to 4 GHz + 3 dB to 70 dB <0.2 dB <1°
70 dB to 80 dB (ZVCE) <1 dB <6°
70 dB to 85 dB (ZVC) <1 dB <6°
4 GHz to 8 GHz +3 dB to -35 dB <0.2 dB <2°
35 dB to 45 dB (ZVCE) <1 dB <6°
35 dB to 50 dB (ZVC) <1 dB <6°
Note: Obtaining the linearity is made easier by using a set of Excel files available from Rohde &
Schwarz. If the measured values are entered in the tables of these files, the linearity
deviation is calculated automatically. Please consult our central service department
(5SMF).
5.5.2.3 Noise Level
Test equipment: Test cable ZV-Z21, attenuator RSM
Calibration: Connect PORT1 to PORT2 via test cable and attenuator
Attenuator 0 dB, SOURCE POWER -10 dBm
MEAS : INPUT b1 (INPUT b2)
TRACE : DATA TO MEMORY : SHOW MATH
Test setup: Connect PORT1 to PORT2 via test cable and attenuator. Attenuator 110 dB.
Settings at the network analyzer:
PRESET
SOURCE POWER -25 dBm
MEAS INPUT b1, INPUT b2
IF BANDWIDTH 3 kHz
AVG SWEEP AVG 10
MARKER MARKER CONT
MARKER Marker frequency = test frequency