Test Sequence (R&S ZVC, R&S ZVCE) ZVx
1043.0009.50 5.46 E-7
5.5.1.7 Level Linearity
Test equipment: Power meter NRVD with power sensor NRV-Z5
Test setup:
Connect power sensor to PORT1 or PORT2 of the network analyzer.
Settings at the network analyzer:
SWEEP SINGLE POINT
CENTER Test frequency
SOURCE POWER Maximum value to minimum value
SWEEP TIME 200 s
MEAS INPUT a1
INPUT a2
Measurement:
Set test frequencies according to test report, read the level on the power meter.
Permissible deviations referred to the value at -10 dBm:
f = 40 kHz to 8 GHz
+10 to –5 dB < 0.4 dB
–5 dB to –15 dB < 0.8 dB
With option ZVR-B10 referred to the value at 0 dBm on PORT 1:
Var. 51, 52, 61, 62
20 kHz to 8 GHz
+6 dB to -9 dB < 0.4 dB
-9 dB to -17 dB < 0.6 dB
Var. 50, 60
20 kHz to 8 GHz
+10 dB to -5 dB < 0.4 dB
-5 dB to -15 dB < 0.6 dB