Specifications
Test Sequence (R&S ZVC, R&S ZVCE) ZVx
1043.0009.50 5.42 E-7
5.5 Test Sequence (R&S ZVC, R&S ZVCE)
The rated specifications of the network analyzer are checked after a warm-up time of at least 30
minutes and after performing the reference oscillator adjustment, recording of detector characteristics
and factory calibration. This makes sure that the guaranteed data are observed. The device settings
quoted are PRESET settings.
The values given in the following sections are not guaranteed, only the technical data of the data sheet
are binding.
5.5.1 Testing the Generator Specifications
5.5.1.1 Frequency Accuracy
Test equipment: R&S FMB with Option R&S FMA-B10
Test setup: Connect FMB (operating mode COUNTER) to PORT1 of the network analyzer.
Settings at the network analyzer:
PRESET
CENTER Test frequency
SWEEP SINGLE POINT
SWEEP TIME 500 s
SOURCE POWER Maximum level
MEAS INPUT a1
Measurement:
Set test frequencies according to test report.
Deviation: Frequency value displayed on the FMB minus
setting value.
Permissible deviation < 2ppm + 1ppm/year
5.5.1.2 Harmonics Suppression
Test equipment: Test cable, option R&S ZVR-B4, attenuator RSG or RSM
Test setup: Connect RSG (RSM) via test cable between PORT1 and PORT2.
Settings of the attenuator: 30 dB
Settings at the network analyzer:
PRESET
MODE FREQUENCY CONVERSION : SECOND
HARMONIC, THIRD HARMONIC
MARKER Test frequency
SOURCE POWER Pmax and –10 dBm (Pmax and 0 dBm with option
R&S ZVR-B10)
MEAS INPUT b2, DRIVE PORT 1