Specifications

Test Sequence (R&S ZVR, R&S ZVRE, R&S ZVRL) ZVx
1043.0009.50 5.16 E-7
5.2.3.4 Crosstalk
Test equipment: 2 N-short-circuits (SHORT male and SHORT female with THRU male from
calibration kit ZV-Z21 or ZV-Z22)
Test setup: Connect N-short-circuits to PORT1 and PORT2.
Settings at the network analyzer:
PRESET
SWEEP LOG SWEEP
SOURCE POWER Maximum value
IF BANDWIDTH 1 Hz
AVERAGE POINT AVG 5
MEAS RATIO b1/b2, DRIVE PORT 2 (crosstalk port 1,
R&S ZVR, R&S ZVRE)
RATIO b2/b1, DRIVE PORT 1 (crosstalk port 2,
R&S ZVR, R&S ZVRE, R&S ZVRL)
MARKER Marker frequency = test frequency
Measurement: Read out marker value on the network analyzer.
Permissible crosstalk values:
R&S ZVR R&S ZVRE, R&S ZVRL
Test Set 50
20 kHz to 200 kHz –90 dB 90 dB
200 kHz to 5 MHz –120 dB –120 dB
5MHz to 1 GHz –130 dB –125 dB
1 GHz to 3 GHz –120 dB –115 dB
3 GHz to 4 GHz –110 dB –105 dB
Test Set 75
20 kHz to 200 kHz –84 dB 84 dB
200 kHz to 5 MHz –114 dB –114 dB
5MHz to 1 GHz –124 dB –119 dB
1 GHz to 3 GHz –114 dB –109 dB
3 GHz to 4 GHz –104 dB –99 dB
R&S ZVR, R&S ZVRE with Option R&S ZVR-B10
20 kHz to 200 kHz –75 dB
200 kHz to 5 MHz –100 dB
5MHz to 1 GHz –110 dB
1 GHz to 3 GHz –100 dB
3 GHz to 4 GHz –90 dB