Specifications

ZVx Test Sequence (R&S ZVR, R&S ZVRE, R&S ZVRL)
1043.0009.50 5.13 E-7
5.2.3 Testing the Test Set Specifications
5.2.3.1 Matching to PORT1 and PORT2
R&S ZVR, R&S ZVRE
Test equipment
test set 50 :
Test equipment
test set 75 :
Calibration Kit ZV-Z21, Test Port Cable pair ZV-Z11
Calibration Kit ZV-Z22, Test Port Cable pair ZV-Z12
Test setup: Test cable between PORT1 and PORT2.
Settings at the R&S ZVR or R&S ZVRE, respectively:
PRESET
MEAS S22 for matching PORT1
S11 for matching PORT2
MARKER test frequency
Reference
measurement:
Perform one-port calibration at the end of the cable over the complete frequency
range (measurement PORT1: loosen cable from PORT1 and calibrate;
measurement PORT2: loosen cable from PORT2 and calibrate).
Measurement:
Set marker to test frequencies according to test report.
Record return loss values.
Passive test set:
Test frequency: Return loss
50: 75:
40 kHz to 100 kHz > 10 dB > 6 dB
100 kHz to 100 MHz > 16 dB > 12 dB
100 MHz to 3000 MHz > 18 dB > 18 dB
3000 MHz to 4000 MHz > 16 dB > 15 dB
Active test set:
Test frequency: Return loss
50: 75:
300 kHz to 1 MHz > 6 dB > 4 dB
1 MHz to 100 MHz > 16 dB > 12 dB
100 MHz to 3000 MHz > 18 dB > 18 dB
3000 MHz to 4000 MHz > 16 dB > 10 dB