Specifications
Test Sequence (R&S ZVR, R&S ZVRE, R&S ZVRL) ZVx
1043.0009.50 5.12 E-7
Alternative Measurement (firmware version 1.50 and higher):
Test equipment,
test set 50 :
Test equipment,
test set 75 :
Calibration kit ZV-Z21, pair of test cables ZV-Z11.
Calibration kit ZV-Z21, pair of test cables ZV-Z11, matching pad RAM, calibration kit
ZCAN-75.
Test setup:
Connect test cable between PORT2 and Input b1 or between PORT1 and Input b2
(for 75 : screw 75 through male, RAM and 50 through female to PORT2 or
PORT1)
Settings at the network analyzer:
PRESET
SWEEP LOG SWEEP
MARKER Test frequency
MEAS S22 for measurement Input b1
S11 for measurement Input b2
SERVICE FUNCTION 2.13.1.2.2 for measurement Input b1 (reset: 2.13.0)
2.13.1.4.2 for measurement Input b2 (reset: 2.13.0)
Calibration: Perform one-port calibration at the end of the test cable (towards Input b1 or Input
b2).
Measurement:
Set test frequencies according to test report, measure b1 and b2.
Take return loss values.
Test frequency
1)
: Return loss:
300 kHz to 4 GHz > 8 dB
1) Measurement below 300 kHz not required because the matching value is uniquely determined by the design at low
frequencies. A possible fault is recognized when the level accuracy is measured.