Specifications
ZVx Test Sequence (R&S ZVR, R&S ZVRE, R&S ZVRL)
1043.0009.50 5.9 E-7
Measurement:
Set test frequencies according to test report.
Read out marker value on the ZVx
Permissible deviation from the applied nominal level (–10 dBm) at PORT1 or PORT2,
respectively:
Passive test set: Active test set:
Frequency range
R&S ZVR/E/L 50/75
Frequency range
R&S ZVR/E/L 50/75
9 kHz to 100 kHz +2 dB
–2 dB
300 kHz to 1 MHz +2 dB
–2 dB
100 kHz to 4 GHz +1 dB
–1 dB
1 MHz to 4 GHz +1 dB
–1 dB
Measurement
R&S ZVR-B25:
Permissible deviation from the applied nominal level (–10 dBm) at Input b1 or Input
b2:
10Hz to 100 kHz +1 dB/ -1 dB
100 kHz to 100 MHz +1 dB/ -1 dB
100 MHz to 4 GHz +1 dB/ -1 dB
5.2.2.2 Linearity
Test equipment:
Attenuator RSG (alternatively one 10 dB plus two 20 dB attenuation pads), 2
matching pads RAM for 75- test set.
Test setup: Connect RSG between PORT1 and PORT2.
Settings at the network analyzer:
PRESET
CAL POWER UNCAL off
MODE SWEEP MODE : POWER SWEEP
SWEEP NUMBER OF POINTS = 51
SOURCE Test frequency
START -25 dBm
STOP Max
SWEEP TIME Auto
IF BANDWIDTH 10 Hz
MEAS S12 (Lin. PORT1), S21 (Lin. PORT2)
FORMAT MAGNITUDE,PHASE
MARKER MARKER CONT
MARKER1 = -10 dBm
DELTA MARKER REF MARKER1
MARKER2 Test level
Reference
measurement:
Set RSG to 20 dB (for 75 to 10 dB), set test frequencies and levels according to
the performance test report and record measurement results.