Specifications
ZVx Test Sequence (R&S ZVR, R&S ZVRE, R&S ZVRL)
1043.0009.50 5.5 E-7
5.2.1.5 Residual FM
Test equipment: Modulation meter FMB, matching pad RAM for 75 test set
Test setup: Connect modulation meter (operating mode DEMOD FM DET RMS 10 Hz to
3 kHz) to PORT1 of the network analyzer (RAM to FMB for 75 ).
Settings at the network analyzer:
PRESET
SWEEP SINGLE POINT
CENTER Test frequency
SOURCE POWER Maximum level
SWEEP TIME 200s
MEAS INPUT a1
Measurement Set test frequencies according to test report, read the residual FM values on the
modulation meter.
Permissible residual FM:
9 kHz to 10 MHz <1 Hz
10 MHz to 150 MHz <2 Hz
150 MHz to 1 GHz <5 Hz
1 GHz to 2 GHz <10 Hz
2 GHz to 4 GHz <20 Hz
5.2.1.6 Level Accuracy
Test equipment: Power meter NRVD with power sensor NRV-Z51 (50 ) or NRV-Z51 with matching
pad RAM (75 ).
Test setup 50 :
Test setup 75 :
Connect power sensor to PORT1, PORT2 (only R&S ZVR and R&S ZVRE) or
OUTPUTa1 (only with option R&S ZVR-B25, Ext. Measurements) of the network
analyzer.
Connect power sensor with RAM to PORT1 or PORT2 (only R&S ZVR and R&S
ZVRE) of the network analyzer, Option R&S ZVR-B25, External Measurements,
as with 50 .
Settings at the network analyzer:
PRESET
SWEEP SINGLE POINT
CENTER Test frequency
SOURCE POWER -10 dBm on PORT 1 and PORT 2
+3 dBm with option R&S ZVR-B10 on PORT 1
SWEEP TIME 200 s
MEAS INPUT a1 (PORT1), INPUT a2 (PORT2)