Specifications
TRIGger-Subsystem ZVx
1043.0009.50 3.144 E-15
3.6.19 TRIGger-Subsystem
The TRIGger subsystem is used to synchronize tasks performed by the instrument and events. This
allows to control and synchronize the start of a sweep. An external trigger signal can be applied to the
rear panel connector.
COMMAND PARAMETER UNIT COMMENT
TRIGger
[:SEQuence]
:SOURce
:TIMer
:RTCLock
:HOLDoff
:SLOPe
:LINK
IMMediate | EXTernal | LINE | TIMer |
MANual | RTCLock
<numeric_value>
0...23, 0...59, 0...59
<numeric_value>
POSitive | NEGative
<string>
S
--
S
--
--
TRIGger[:SEQuence]:SOURce
This command selects the trigger source for starting a sweep..
Syntax:
TRIGger[:SEQuence]:SOURce IMMediate | EXTernal | LINE | TIMer |
MANual | RTCLock
Example:
"TRIG:SOUR EXT"
Features:
*RST value:
SCPI:
IMMediate
conforming
TRIGger[:SEQuence]:TIMer
This command selects the internal trigger with a specified time interval.
Syntax:
TRIGger[:SEQuence]:TIMer <numeric value>
<numeric_value> ::= 0 .. 1000s
Example:
"TRIG:TIM 2s"
Features:
*RST value:
SCPI:
-
conforming
TRIGger[:SEQuence]:RTCLock
This command selects the internal clock with a specified time as a trigger signal.
Syntax:
TRIGger[:SEQuence]:RTCLock 0...23, 0...59, 0...59
Example:
"TRIG:RTCL 13,30,0"
Features:
*RST value:
SCPI:
-
device-specific