Specifications

SOURce Subsystem ZVx
1043.0009.50 3.120 E-15
SOURce<1...4>: POWer:CORRection:EXT<1|2>:SWEep
This command defines the sweep parameters to be used when an external generator power
calibration is started.
Syntax:
SOURce<1...4>:POWer:
CORRection:EXT<1|2>:SWEep
<numeric_value>,<numeric_value>
,<numeric_value>, LIN | LOG
Example:
"SOUR:POW:CORR:EXT1:SWE 10MHZ,1GHZ,101,LOG"
Features:
*RST value:
SCPI:
.., .., 101, LIN (Start / Stop depending on external generator)
conforming
The parameters are sorted as follows:
start frequency (range depending on external generator)
stop frequency (range depending on external generator)
number of frequency points (1 .. 2001)
sweep mode (LIN, LOG)
If the number of points is set equal to 1, the start and stop frequencies must be the same
SOURce<1...4>: POWer:CORRection:NREadings
This command defines the number of power measurements at one frequency point when a power
calibration is started.
Syntax:
SOURce<1...4>:POWer:CORRection:NREadings <numeric_value>
<numeric value> ::= 1 .. 100
Example:
"SOUR:POW:CORR:NRE 2"
Features:
*RST value:
SCPI:
1
conforming
SOURce<1...4>: POWer:CORRection:[:ACQuire]
This command is used to perform a generator power calibration (internal, external).
Syntax:
SOURce<1...4>:POWer:CORRection[:ACQuire] A1 | A2 | ESRC1 | ESRC2
Example:
"SOUR:POW:CORR:ACQ A1"
Features:
*RST value:
SCPI: conforming
SOURce<1...4>: POWer:CORRection:LLISt
This command defines the power loss list which may be used for a power calibration.
Syntax:
SOURce<1...4>:POWer:CORRection:LLISt <numeric_value>,
<numeric_value>,
<numeric_value> ...
Example:
"SOUR:POW:CORR:LLIS 2,10HZ,1DB,4GHZ,2DB"
Features:
*RST value:
SCPI:
0 (i.e. power loss list empty)
conforming
The parameters are sorted as follows:
number of following (frequency, attenuation) pairs (0 to 20)
frequency 1 (10 Hz to 4 GHz)
attenuation for frequency 1 (-100 dB to 1000 dB)
...
The frequencies must be sorted in ascending order.