Specifications

SOURce Subsystem ZVx
1043.0009.50 3.118 E-15
SOURce[1...4]:POWer:NLINear:SOI:RANGe:LOWer
This command defines a lower limit for the power of the signal sources for a 2
nd
order intercept point
measurement. The allowed range of lower limits depends on the power of the signal sources.
With power calibration, the lower limit must be set such that it can be reached by both signal
sources.
Syntax:
SOURce[1...4]: POWer:NLINear:SOI:RANGe:LOWer <numeric_value>
Example:
"SOUR:POW:NLIN:SOI:RANG:LOW -25dBm"
Features:
*RST value:
SCPI:
-25 dBm
device-specific
SOURce[1...4]:POWer:NLINear:TOI:RANGe:UPPer
This command defines an upper limit for the power of the signal sources for a 3
rd
order intercept
point measurement. The allowed range of upper limits depends on the power of the signal sources.
With power calibration, the upper limit must be set such that it can be reached by both signal
sources.
Syntax:
SOURce[1...4]: POWer:NLINear:TOI:RANGe:UPPer <numeric_value>
Example:
"SOUR:POW:NLIN:TOI:RANG:UPP 0dBm"
Features:
*RST value:
SCPI:
0 dBm
device-specific
SOURce[1...4]:POWer:NLINear:TOI:RANGe:LOWer
This command defines a lower limit for the power of the signal sources for a 3
rd
order intercept point
measurement. The allowed range of lower limits depends on the power of the signal sources.
With power calibration, the lower limit must be set such that it can be reached by both signal
sources.
Syntax:
SOURce[1...4]: POWer:NLINear:TOI:RANGe:LOWer <numeric_value>
Example:
"SOUR:POW:NLIN:TOI:RANG:LOW -25dBm"
Features:
*RST value:
SCPI:
-25 dBm
device-specific
SOURce<1...4>: POWer:CENTer
This command defines the center level of the analyzer in the level sweep mode.
Syntax:
SOURce<1...4>: POWer:CENTer <numeric_value>
<numeric value> ::= -25dBm to 0 dBm (device-specific)
Example:
"SOUR:POW:CENT -10dBm"
Features:
*RST value:
SCPI: conforming