ZVx SENSe Subsystem
1043.0009.50 3.105 E-15
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:FUNDamental
For mixer measurements, this command selects the fundamental frequency out of the three mixer
frequencies. .
Syntax:
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:FUNDamental RF | LO | IF
Example:
"FREQ:CONV:MIX:FUND RF"
Features:
*RST value:
SCPI:
LO
device-specific
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:LOEXternal
This command selects one of the two external signal sources as a local oscillator.
Syntax:
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:LOEXternal SOURCE1 |
SOURCE2
Example:
"FREQ:CONV:MIX:LOEX SOURCE2"
Features:
*RST value:
SCPI:
SOURCE1
device-specific
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:RFFixed
This command defines the RF mixer frequency as a fixed value.
Syntax:
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:RFFixed <numeric_value>
Example:
"FREQ:CONV:MIX:RFF 2.5GHz"
Features:
*RST value:
SCPI:
0
device-specific
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:LOFixed
This command defines the LO mixer frequency as a fixed value.
Syntax:
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:LOFixed <numeric_value>
Example:
"FREQ:CONV:MIX:LOF 1.8GHz"
Features:
*RST value:
SCPI:
0
device-specific
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:IFFixed
This command defines the IF mixer frequency as a fixed value.
Syntax:
[SENSe[1...4]:]FREQuency:CONVersion:MIXer:IFFixed <numeric_value>
Example:
"FREQ:CONV:MIX:IFF 21.4MHz"
Features:
*RST value:
SCPI:
0
device-specific