Specifications

ZVx SENSe Subsystem
1043.0009.50 3.99 E-15
The table below shows the correction terms available for the calibration methods:
Calibration Method Available Correction Terms
Trans Norm Forward
Trans Norm Reverse
SCORR6
SCORR12
Trans Norm both Directions SCORR6, SCORR12
Refl Norm P1
Refl Norm P2
SCORR3
SCORR9
Refl Norm both Ports SCORR3, SCORR9
Trans+Refl Norm Forward
Trans+Refl Norm Reverse
SCORR3, SCORR6
SCORR9, SCORR12
Trans+Refl Norm both Ports SCORR3, SCORR6, SCORR9, SCORR12
Full One Port P1
Full One Port P2
SCORR1 to SCORR3
SCORR7 to SCORR9
Full One Port both Ports SCORR1 to SCORR3, SCORR7 to SCORR9
One Path Two Port Forward
One Path Two Port Reverse
SCORR1 to SCORR3, SCORR6
SCORR7 to SCORR9, SCORR12
TOM, TRM, TNA, TRL G11 to G22 and H11, H12, H22 (H21 = 1);
Gxx is normalized to H21
TOSM SCORR1 to SCORR12
TOM-X
E11 to E22, G11 to G22, H11 to H22, F11, F21, F12
[SENSe[1...4]:]CORRection:DATE?
This command queries the date at which the active system error correction was recorded.
Syntax:
[SENSe[1...4]:]CORRection:DATE?
Example:
"CORR:DATE?"
Features:
*RST value:
SCPI:
-
device-specific
[SENSe[1...4]:]CORRection:POWer:ACQuire
This command is used to acquire a receiver power calibration.
Syntax:
[SENSe[1...4]:]CORRection:POWer:ACQuire B1 | B2 | IFRef
Example:
"CORR:POW:ACQ B1"
Features:
*RST value:
SCPI: conforming
[SENSe[1...4]:]CORRection:POWer[:STATe]
This command switches on or off the power calibration in the specified channel.
Syntax:
[SENSe[1...4]:]CORRection:POWer[:STATe]
Example:
"CORR:POW:STAT ON"
Features:
*RST value:
SCPI:
OFF
conforming