Specifications

ZVx SENSe Subsystem
1043.0009.50 3.97 E-15
[SENSe[1...4]:]CORRection:EDELay[1|2]:ELENgth
This command defines the length offset as an electrical length.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2]:ELENgth <numeric value>
Example:
"CORR:EDEL:ELEN 12.32mm"
Features:
*RST value:
SCPI:
0
device-specific
[SENSe[1...4]:]CORRection:EDELay[1|2]:DIELectric
This command defines the value of the dielectric corresponding to the length specified via
EDELay:DISTance.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2]:DIELectric <numeric value>
Example:
"CORR:EDEL2:DIEL 1.2"
Features:
*RST value:
SCPI:
1
device-specific
[SENSe[1...4]:]CORRection:EDELay[1|2]:AUTO
This command sets the length offset such that the phase offset of the quantity measured in the
active channel is minimized over the current sweep.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2]:AUTO ONCE
Example:
"CORR:EDEL2:AUTO ONCE"
Features:
*RST value:
SCPI:
device-specific
This command is an event, which is why it is not assigned an *RST value and has no query.
[SENSe[1...4]:]CORRection:EDELay[1|2]:STATe
This command switches the correction of the length offset on or off.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2]:State ON | OFF
Example:
"CORR:EDEL:STAT ON"
Features:
*RST value:
SCPI:
OFF
conforming
[SENSe[1...4]:]CORRection:OFFSet[1|2]:STATe
This command switches the correction of the amplitude, phase and length offset on and off.
Syntax:
[SENSe[1...4]:]CORRection:OFFSet[1|2]:STATe ON | OFF
Example:
"CORR:OFFS:STAT ON"
Features:
*RST value:
SCPI:
OFF