Specifications
SENSe Subsystem ZVx
1043.0009.50 3.96 E-15
Two-port standards are either MM = connector/connector, FF = socket/socket or MF =
connector/socket, one-port standards are either M = connector or F = socket.
The following standards are available:
THRough Through connection
LINE1 Line 1 for TRL-procedure (ZVR and ZVC only)
LINE2 Line 2 for TRL-procedure (ZVR and ZVC only)
ATTenuation matched attenuator pad (ZVR and ZVC only)
SNETwork reflection symmetric network (ZVR and ZVC only)
OPEN open
SHORt short circuit
REFLect unknown one-port standard (ZVR and ZVC only)
MATCh broadband termination
SMATch sliding load
In case of a query, the first parameter must be specified, not the second.
[SENSe[1...4]:]CORRection:CKIT:SMA
This command selects a standard from the SMA-calibration set.
Syntax:
[SENSe[1...4]:]CORRection:CKIT:SMA
Example:
"CORR:CKIT:SMA MFLINE1"
Features:
*RST value:
SCPI:
-
device-specific
The definition of the abbreviations is analogous to the definition in the N-calibration set. In case of a
query, the first parameter must be specified, not the second.
[SENSe[1...4]:]CORRection:CKIT:PC[7|35]
This command selects a standard from the PC7-, PC3.5 or 2.92 mm-calibration set.
Syntax:
[SENSe[1..4]:]CORRection:CKIT:
Example:
"CORR:CKIT:PC35 FMAT"
Features:
*RST value:
SCPI:
-
device-specific
The definition of the abbreviations is analogous to the definition in the N-calibration set. In case of a
query, the first parameter must be specified, not the second.
[SENSe[1...4]:]CORRection:EDELay[1|2][:TIME]
This command defines the length offset as a delay time.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2][:TIME] <numeric value>
Example:
"CORR:EDEL2 10ns"
Features:
*RST value:
SCPI:
0
conforming
[SENSe[1...4]:]CORRection:EDELay[1|2]:DISTance
This command defines the length offset as a mechanical length.
Syntax:
[SENSe[1...4]:]CORRection:EDELay[1|2]:DISTance <numeric value>
Example:
"CORR:EDEL:DIST 10mm"
Features:
*RST value:
SCPI:
0
conforming