Specifications
ZVx SENSe Subsystem
1043.0009.50 3.93 E-15
[SENSe[1...4]:]CORRection:COLLect:METHod
This command defines the calibration method.
Syntax:
[SENSe[1...4]:]CORRection:COLLect:METHod FTRans | RTRans | FRTRans |
TOM | TRM | TRL | TNA |
TOMX| TOSM | FUNDamental |
FOPORT1 | FOPORT2 |
FOPORT12 | FOPTport |
ROPTport | REFL1 | REFL2 |
REFL12 | TPORT | FTREF1 |
RTREF2
Example:
"CORR:COLL:METH TOM"
Features:
*RST value:
SCPI:
–
conforming
[SENSe[1...4]:]CORRection:COLLect:SAVE
This command calculates the calibration data according to the calibration method currently selected
from the previously performed measurements of the standards and stores them internally.
Syntax:
[SENSe[1...4]:]CORRection:COLLect:SAVE
Example:
"CORR:COLL:SAV"
Features:
*RST value:
SCPI:
–
conforming
This command is an event,, which is why it is not assigned an *RST value and has no query.
[SENSe[1...4]:]CORRection:COLLect:CONNection[1|2]
This command selects a connector family for the port between connector and socket.
Syntax:
[SENSe[1...4]:]CORRection
:COLLect:CONNection[1|2] N50FEMALE | N50MALE |
N75FEMALE | N75MALE | PC7 |
SMAFEMALE | SMAMALE |
PC35FEMALE | PC35MALE |
PC292FEMALE | PC292MALE |
UFEMALE1 | UMALE1 |
UFEMALE2 | UMALE2
Example:
"CORR:COLL:CONN2 SMAM"
Features:
*RST value:
SCPI:
–
device-specific
[SENSe[1...4]:]CORRection:CKIT
This command determines the connector system of the calibration standards.
Syntax:
[SENSe[1...4]:]CORRection:CKIT N50 | N75 | SMA | PC7 | PC35 | PC292 |
USER1 | USER2
Example:
"CORR:CKIT SMA"
Features:
*RST value:
SCPI:
–
device-specific