Specifications

SENSe Subsystem ZVx
1043.0009.50 3.76 E-15
3.6.14 SENSe Subsystem
The SENSe subsystem is divided up into several subsystems. The commands of these subsystems
directly control device-specific settings, they do not refer to the signal characteristics of the
measurement signal.
3.6.14.1 SENSe:AVERage Subsystem
The SENSe:AVERage subsystem calculates the average of the data acquired: Various successive
measurements are combined in order to obtain a new test result. The number of test points and the
abscissa of the new result correspond to those of the original measurements.
COMMAND PARAMETER UNIT COMMENT
[:SENSe[1 to 4]]
:AVERage
: COUNt <numeric_value> --
[:STATe] <Boolean> --
:CLEar -- -- no query
:MODE SWEep| POINt --
[:SENSe[1 to 4]:]AVERage:COUNt
The command specifies the number of measurements combined.
Syntax: [:SENSe[1 to 4]:]AVERage:COUNt <numeric_value>
<numeric value> ::= 0 to 256
Example: "AVER:COUN 16"
Features: *RST value: 0
SCPI: conforming
[:SENSe[1 to 4]:]AVERage[:STATe]
The command switches on or off the average function.
Syntax: [:SENSe[1 to 4]:]AVERage[:STATe] ON | OFF
Example: "AVER OFF"
Features: *RST value: OFF
SCPI: conforming
[:SENSe[1 to 4]:]AVERage:CLEar
The command resets the result of the average function and restarts the measurement.
Syntax: [:SENSe[1 to 4]:]AVERage:CLEar
Example: "AVER:CLE"
Features: *RST value: -
SCPI: device-specific
This command is an event, which is why it is not assigned an *RST value and has no query.