Specifications

ZVx OUTPut Subsystem
1043.0009.50 3.73 E-15
OUTPut:UPORt<1|2>[:VALue]
This command sets the control lines of the user port. If the user port was set to INPut previously, the
output value is stored intermediately.
Syntax: OUTPut:UPORt<1|2>[:VALue] <Binary>
<Binary>::= 00000000 ... 11111111
Example: "OUTP:UPOR2 #B10100101"
Features: *RST value:
SCPI: device-specific
OUTPut:UPORt<1|2>:STATe
This command toggles the control lines of the user ports between INPut and OUTPut.
Syntax: OUTPut:UPORt<1|2>:STATe ON | OFF
Example: "OUTP:UPOR:STAT ON"
Features: *RST value: OFF
SCPI: device-specific
ON switches the user port to OUTPut, OFF switches to INPut.