Specifications

OUTPut Subsystem ZVx
1043.0009.50 3.72 E-15
3.6.12 OUTPut Subsystem
The OUTPut subsystem controls the output features of the analyzer.
COMMAND PARAMETER UNIT COMMENT
OUTPut<1|2>
:ATTenuation
:DPORt
:POWer
:RMIXer
[:STATe]
:UPORt<1|2>
[:VALue]
:STATe
<numeric_value>
PORT1 | PORT2
NORMal | HIGH
<Boolean>
<Binary>
<Boolean>
DB
OUTPut[1|2]:ATTenuation
This command determines the attenuation of the attenuator located in the signal path of the wave a1
or a2 propagating towards the DUT.
Syntax: OUTPut:ATTenuation <numeric_value>
<numeric_value>::= 0dB .. 70dB
Example: "OUTP:ATT 40dB"
Features: *RST value:
SCPI: conforming
OUTPut:DPORt
This command defines the setting of the signal path switch in the test set (drive port).
Syntax:
OUTPut:DPORt PORT1 | PORT2
Example:
"OUTP:DPOR PORT2"
Features:
*RST value:
SCPI:
PORT1
device-specific
OUTPut[1|2]:POWer
This command switches an additional attenuator on and off in the generator signal path. This can
increase the output power to the detriment of matching.
Syntax:
OUTPut[1|2]:POWer NORMal | HIGH
Example:
"OUTP:POW HIGH"
Features:
*RST value:
SCPI:
NORM
device-specific
OUTPut:RMIXer
This command defines whether the reference wave a1 is to be applied internally or via rear-panel
connectors a1 EXT OUT and a1 EXT IN.
Syntax:
OUTPut:RMIXer[:STATe] ON | OFF
Example:
"OUTP:RMIX ON"
Features:
*RST value:
SCPI:
OFF
device-specific