Technical data
1-2
Specifications and Regulatory Information
Specifications and Characteristics
Specifications and Characteristics
Specifications apply to instruments in the following situation:
• temperature is in the range of +20°C to +30°C
• analyzer has had a warm-up time of two hours in a stable ambient temperature
• measurement calibration has been performed
Performance Definitions
Specifications: Warranted performance. Specifications include guardbands to account for the expected
statistical performance distribution, measurement uncertainties, and changes in performance due to
environmental conditions.
Characteristics: Useful, non warranted, information about the functions and performance of the system.
Calibration Cycle
Agilent Technologies warrants instrument specifications over the recommended calibration interval. To
maintain specifications, periodic recalibrations are necessary. We recommend that the analyzer be cali-
brated at an Agilent Technologies service facility every 12 months.
User Calibration Cycle
A user calibration, also known as a measurement calibration, should be performed at least once every 8
hours. If the ambient temperature drifts, then you should perform a calibration more frequently.