Technical data
1-14
Specifications and Regulatory Information
Specifications and Characteristics
Table 1-4. General Information
General Information
Description Specification Characteristic
General Environmental
RFI/EMI Susceptibility Defined by CISPR Pub. 11 and FCC Class B
standards.
ESD Minimize using static-safe work procedures and
an antistatic bench mat
(part number 9300-0797).
Dust Minimize for optimum reliability.
Operating Environment
Temperature +20 °C to +30 °C Instrument powers up, phase locks, and displays
no error messages within this temperature
range.
Humidity 5% to 95% at +30
°C (non-condensing)
Altitude 0 to 4.5 km (15,000 ft)
Storage Conditions
Temperature
−40 ×°C to +55 °C
Humidity 5% to 95% RH at +40
°C
(non-condensing)
Altitude 0 to 15.24 km (50,000 ft)
Cabinet Dimensions
Height x Width x Depth (323 x 430x 476 mm)
(12.71 x 16.93 x 18.74 inches)
Cabinet dimensions exclude front and rear
protrusions.
Weight
Shipping 151 lb
Net 76 lb
Internal Memory - Data Retention Time with 3 V, 1.2 Ah Battery
a
a. Analyzer power is switched off.
70 °C 250 days (0.68 year)
40
°C 1244 days (3.4 years)
25