Error Messages
Error Messages in Alphabetical Order
instrument commands.
SELF TEST #n FAILED. Service Error Number 112. Internal test #n has failed. Several internal test routines are
executed at instrument preset. The analyzer reports the first failure detected.
SEQUENCE ABORTED. Error Number 157. The sequence running was stopped prematurely when you pressed
the Local, key.
SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE. Error Number 153. When you pause a sequence, you
cannot continue it if you have modified it. You must start the sequence again.
SLIDES ABORTED (MEMORY REALLOCATION). Error Number 73. You cannot perform sliding load
measurements due to insufficient memory. Increase the available memory by clearing one or more
save/recall registers and pressing Preset, or by storing files to a disk and then deleting them from internal
memory.
SOURCE POWER DISABLED, EDIT LIST MODE TBL. Information Message. When list power has been enabled
and swept list mode is on, you will not be able to change the power level using the Power, key. To change
the power level, edit the swept list table.
SOURCE POWER TURNED OFF, RESET UNDER POWER MENU. Information Message. You have exceeded the
maximum power level at one of the inputs and power has been automatically reduced. The annotation P?
indicates that power trip has been activated. When this occurs, reset the power and then press Power,
SOURCE PWR on OFF, to switch on the power.
STARTING COPY SPOOLER. Information Message. The analyzer is beginning to output data from the spool
buffer to the copy device. The analyzer resumes normal operation; the data is being output to the copy
device in the background.
SWEEP MODE CHANGED TO CW TIME SWEEP. Error Number 187. If you select external source auto or manual
instrument mode and you do not also select CW mode, the analyzer is automatically switched to CW.
SWEEP TIME INCREASED. Error Number 11. You have made instrument changes that cause the analyzer
sweep time to be automatically increased. Some parameter changes that cause an increase in sweep time
are narrower IF bandwidth, an increase in the number of points, and a change in sweep type.
SWEEP TIME TOO FAST. Error Number 12. The fractional-N and digital IF circuits have lost synchronization.
SWEEP TRIGGER SET TO HOLD. Information Message. The instrument is in a hold state and is no longer
sweeping. To take a new sweep, press Sweep Setup, TRIGGER MENU, SINGLE, or CONTINUOUS.
SYNTAX ERROR. Error Number 33. You have improperly formatted a GPIB command. Refer to the
programmer’s guide for proper command syntax.
SYST CTRL OR PASS CTRL IN LOCAL MENU. Error Number 36. The analyzer is in talker/listener mode. In this
mode, the analyzer cannot control a peripheral device on the bus. Use the local menu to change to system
controller or pass control mode.
TEST PORT OVERLOAD, REDUCE POWER. Error Number 57. You have exceeded approximately +14 dBm at one
of the test ports (or 0 dBm at the A or B sampler, Option 012 only). When this occurs, reduce the power to
a lower level.
THIS LIST FREQ INVALID. Error Number 133. You have set frequencies in the list that are outside of the