Technical data
6-11
Error Messages
Error Messages in Alphabetical Order
printer address recognized by the analyzer matches the GPIB address set on the printer itself.
PRINTER: paper error. Error Number 171. There is a paper-related problem with the parallel port printer such
as a paper jam or out-of-paper condition.
PRINTER: power off. Error Number 174. The power to the printer at the parallel port is off.
PRINTER: reset in progress. Information Message. If the printer takes longer than a half-second to reset, this
message will be displayed until printer is finished with reset.
PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL. Error Number 166. If a print or plot is in progress and you
attempt a second print or plot, this message is displayed and the second attempt is ignored. To abort a
print or plot in progress, press Local.
PROCESSING DISPLAY LIST. Information Message. The display information is being processed for a screen
print to a copy device and stored in the copy spool buffer. During this time, the analyzer's resources are
dedicated to this task (which takes less than a few seconds).
PWR MTR NOT ON/CONNECTED OR WRONG ADDRS. Error Number 117. The power meter cannot be accessed
by the analyzer. Verify that the power meter address and model number set in the analyzer match the
address and model number of the actual power meter.
RANGE CAUSED POWER LVL CHANGE IN LIST. Error Number 213. The selected power range changed the
power level of one or more segments in the swept list table. Change the segment power or change the
power range.
RECEIVER TEMPERATURE LOOP OPEN. Error Number 219. The receiver temperature is unstable. (This message
may appear for the first few minutes when the analyzer is powered up.) Cycle the analyzer power and allow
a warm-up time of two hours, then press any key. If the message persists, contact your nearest Agilent
office.
REQUESTED DATA NOT CURRENTLY AVAILABLE. Error Number 30. The analyzer does not currently contain the
data you have requested. For example, this condition occurs when you request error term arrays and no
calibration is active.
RIPPLE LIMIT TABLE EMPTY. Information Message. The ripple limit table does not have any frequency bands
defined. Add at least one frequency band to the ripple limit table for ripple testing.
SAVE FAILED. INSUFFICIENT MEMORY. Error Number 151. You cannot store an instrument state in an internal
register due to insufficient memory. Increase the available memory by clearing one or more save/recall
registers and pressing Preset, or by storing files to a disk.
SEGMENT #n POWER OUTSIDE RANGE LIMIT. Information Message. The selected power range does not
support the power level of one or more segments in the swept list table. This message appears when swept
list mode is not on and reports the first segment that is out of range. Change the segment power or change
the power range.
SEGMENT #n START FREQ OVERLAPS PREVIOUS SEGMENT. Information Message. A segment entered in the
swept list table caused one or more frequency segments to overlap. This message appears when swept list
mode is not on and reports the first segment that is overlapping another. Change the frequency ranges of
the overlapping segments.
SELECTED SEQUENCE IS EMPTY. Error Number 124. The sequence you attempted to run does not contain