Technical data
6-8
Error Messages
Error Messages in Alphabetical Order
LIMIT TABLE EMPTY. Error Number 205. Limit lines cannot be turned on unless a limit table has been created.
Refer to the “Making Measurements” chapter of the user’s guide for information on how to create a limit
table.
LIST TABLE EMPTY. Error Number 9. The frequency list is empty. To implement list frequency mode, add
segments to the list table.
LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN. Error Number 150. A logarithmic sweep is only valid if the
stop frequency is greater than four times the start frequency. For frequency spans of less than two octaves,
the sweep type automatically reverts to linear sweep.
LW TESTSET MISSING. Error Number 224. The lightwave testset may be powered down separately, or the
cable at the rear panel may be removed or loose.
MEMORY FOR CURRENT SEQUENCE IS FULL. Error Number 132. All the memory in the sequence you are
modifying is filled with instrument commands.
MODULATOR TEMPERATURE LOOP OPEN. Error Number 220. The modulator temperature is unstable. (This
message may appear for the first few minutes when the analyzer is powered up.) Cycle the analyzer power
and allow a warm-up time of two hours, then press any key. If the message persists, contact your nearest
Agilent office.
MORE SLIDES NEEDED. Error Number 71. When you use a sliding load (in a user-defined calibration kit), you
must set at least three slide positions to complete the calibration.
NO CALIBRATION CURRENTLY IN PROGRESS. Error Number 69. The RESUME CAL SEQUENCE, softkey is not
valid unless a calibration is already in progress. Start a new calibration.
NO DISK MEDIUM IN DRIVE. Error Number 41. You have no disk in the current disk unit. Insert a disk, or check
the disk unit number stored in the analyzer.
NO FAIL FOUND.Service Error Number 114. The self-diagnose function of the instrument operates on an
internal test failure. At this time, no failure has been detected.
NO FILE(S) FOUND ON DISK. Error Number 45. No files of the type created by an analyzer store operation were
found on the disk or the disk drive is empty. If you requested a specific file title, that file was not found on
the disk.
NO IF FOUND: CHECK R INPUT LEVEL. Error Number 5. The first IF signal was not detected during pretune.
Check the front panel R channel jumper. If there is no visible problem with the jumper, refer to the service
guide for troubleshooting.
NO LIMIT LINES DISPLAYED. Error Number 144. You can turn limit lines on but they cannot be displayed on
polar or Smith chart display formats.
NO MARKER DELTA - SPAN NOT SET. Error Number 15. You must turn the delta marker mode on, with at least
two markers displayed, in order to use the MARKER -> SPAN, softkey function.
NO MEMORY AVAILABLE FOR INTERPOLATION. Error Number 123. You cannot perform interpolated error
correction due to insufficient memory.
NO MEMORY AVAILABLE FOR SEQUENCING. Error Number 126. You cannot modify the sequence due to
insufficient memory.