Technical data
1-4
Specifications and Regulatory Information
Specifications and Characteristics
Optical-to-Optical Device Measurement Specifications
The following data applies after a response and isolation calibration has been performed. Connectors
should be HMS-10 or equivalent.
O/O Noise Floor
Optical-to-Electrical Device Measurement Specifications
Relative frequency response can be used to calculate the error in measuring the 3 dB bandwidth of an O/E
device.
Relative Frequency Response Performance Data
Optical-to-Optical Measurement Performance Data
Description Frequency Range Noise Floor (dBo)
Maximum Noise Floor Amplitude
a
a. Noise Floor is measured with 30 Hz IF bandwidth and with an averaging factor of 6.
0.05 to 8 GHz –30
8 to 20 GHz –25
Optical-to-Electrical Measurement Performance Data
Description Frequency Range
Specification
a
a. Applies to a device with ρ = <0.25 and measurement settings of IF bandwidth = 30 Hz and smoothing = 0.5%.
System Relative Frequency Response Accuracy 0.05 to 11 GHz ±0.65 dB
11 to 20.05 ±0.90 dB