Specifications
F. No. 85/Path/RPC/SSK/12-13
Cryostat – 01 No.
- The Instrument should be free-standing motorized open top with independent specimen temperature control
via Two separate compressor system.
- Convenient section thickness selection from outside the cryo-chamber.
- Section thickness range : 0.5 to 300 um.
- Temperature setting range –10 deg. To –50 deg. C.
- Cryo-chamber temperature setting down upto –40 deg. C.
- Specimen retraction : 50 um.
- Vertical Specimen Stroke : 50 mm
- Programmable Trimming : 5 to 150 um, in steps of 5, 10, 30, 50, 100 &150 um.
- Window setting for cutting length.
- Should have foot switch as well as buttons at control panel for motorized cutting.
- Manual as well as programmable automatic 24 hours defrost cycle.
- Universal Knife holder base, holder for disposable blades, disposable blades.