Specifications

F. No. 76/EB/RPC/SSK/12-13
Class 2 Biosafety Cabinet – 01 No.
The equipment should have following features:
Cabinet should be of Class II type A2, provided with stand.
Normal Size of Biosafety Cabinet: 1.2 meters (4 feet).
External Dimensions: Width= 1300-1350mm, Depth: 700-750mm, Height:1350 - 1400mm
Internal Dimensions = Width: 1250 - 1300mm, Depth: 550-600mm, Height:650- 700 mm
Cabinet should have inflow of 100 feet per minute(fpm) and down flow 70 fpm
Should have Dynamic Chamber plenum design for quiet, uniform airflow.
Microprocessor controlled with two ULPA (Ultra Low Penetration air)filters, efficiency of >99.999% at
0.1 to 0.3µ to provide 70% down flow and 30% exhaust with under pressure condition in the working
area.
Cabinet should have frameless front window with non- reflecting safety glass (should have advanced
laminated glass).
Cabinet should have very low noise level (62 dBA) according to EN 12469 and vibration free.
Fluorescent Light Intensity of cabinet = >1100 lux
Cabinet should have control system with visual and buzzer alarm with cabinet information.
For Air Cleanliness: ISO 14664.1 Class 3 and other equivalent air cleanliness requirements.
Average air flow- down flow at initial set point; 0.30 m/s and inflow 0.45 m/s
Cabinet should have 220 V/50Hz power supply.
Main body should made up 18 gauge electro-galvanized steel, with white oven backed epoxy
antimicrobial powder coated finish.
Biosafety cabinets should have ducting system.
Demonstration, installation and complete training of the biosafety cabinet should be provided.
Biosafety Cabinet should have air quality of biosafety standards and installation should include the
certificate of standard air quality.
Biosafety cabinet should be properly installed in the room without damaging existing walls.