F. No. 44/U-4/RPC/SSK/12-13
Laser Machine for selective Laser Trabeculoplasty – 01 No.
Laser System : Portable Q-Switched, frequency-doubled Nd:YAG;532 nm wavelength
Pulse Duration : 3 Nanoseconds
Repetition Rate : 3 Hz
Energy Settings : 0.30 to 2.0 mJ
Pulse Setting : Single pulse-on-demand
Spot Size : 400 micron
Aiming Beam : Diode laser; coaxial, 635 nm wavelength, variably intensity
Laser Delivery : Clip On Type; compatible with slit lamps
Display : Remote Control. Should show exact no. of Shots and total Energy delivered.
Cooling System : Air convection
Power Requirements : 120-230 VAC, 50/60 Hz, switchable input
Weight : 12 Kg
Optional Accessories : SLT Lens, Remote Control
Slit Lamp : Imported 5 Step Magnification Slit Lamp with Imported Motorized Table
Should be US FDA Approved
Service Support : The Principal Company should provide direct service support or should be available
directly in India.