Specifications
F.No. 14/U-1&2/RPC/SSK/12-13
Endo Laser 532 -3 Nos
1. Should be a diode pumped frequency doubled solid state laser
2. Should have a 532 mm operating wavelength
3. Should have a forced air cooling system
4. Should have a aiming laser wavelength of 635 nm
5. Should have a out put power ranging from 30 mW to 2000 mW (2W)
6. Should have a exposure time of 10ms – 2000 ms and continous wave
7. Should be compatible with Endo laser probes
8. Should have provision to connect two doctors filter
9. Should have provision for remote interlock
10. Should have dual laser ports
11. Should have LED illumination at the ports for visual verification
12. Should have power control in the footswitch
13. Should have Ready – Standby control in the footswitch
14. Should have voice confirmation from the machine
15. Should have power source for LIO on the Console itself
16. Should have electric requirement of 220V
17. Should have LED in the footswitch for visualization in the dark OT.