Specifications
F. No. 134/Radio/RPC/SSK/12-13
Portable ultrasound color Doppler machine – 01 No.
A state of art fully digital, compact portable Colour Doppler Ultrasound Machine is required with following
technical features for use in Dr. R.P. Centre.
1. It should be suitable for complete orbital & eye applications, vascular access and examination of other
small parts. It should allow ophthalmic and general applications both in adults and children & also
suitable for interventions.
2. The equipment must be capable of operating in B-mode.
3. It must support transducers with linear, phased array and curved array formats.
4. The system shall have broadband architecture with an operating frequency of at least 1-14 MHz
5. The system shall have the ability to enhance tissue margins and improve contrast resolution by reducing
artifacts and improving visualization of texture patterns.
6. System should have special needle visualization within the image, please specify the technology.
7. Unit should be able to give very high image quality with advance technologies like Compound Imaging
with at least 5 sights of lines for better contrast resolution, tissue differentiation and edge detection.
8. The system shall process a dynamic range that is at least 165 db. The system must display at a
maximum depth of 35 cm.
9. The system shall provide a backlit keypad for ease of use, with facility to disinfect the keypad of system
must be possible to avoid any cross contamination.
10. The system shall go from the off status to active scanning in less than 20 seconds.
11. The system should be easy to carry, the system shall not weigh more than 5 kg including battery.
12. The system shall have an LCD screen size no smaller than 10” and must have provision to connect a
bigger monitor for use in OT.
13. The system shall have Digital Video Interface (DVI), S-Video, VGA, USB and audio out put.
14. The system shall have the ability to function by AC/DC or battery power with the same degree of
functionality, the battery life (run time) shall be at least 2 (two) hours, this needs to be demonstrated.
15. The system must have archive capability for storage and retrieval of images and clips.
16. The system shall support all DICOM functionality, Storage, Print, Work List and also ready to connect
to PACS.
Transducers to be supplied with system
1. 6-13(+_1) MHz multi-frequency, broadband linear array transducer for ophthalmic and vascular imaging
with less than 40 mm probe foot print.
2.
Convex probe of 2- 5 MHz for general applications.
3. 4 – 8 MHz multi-frequency, broadband phased array small foot print transducer for orbital applications
for use both in children and adults.
All probes must be supplied with biopsy guide attachments.
Essential Accessories : B/W Thermal printer, original carry bag, original moulded
mobile cart (from the principal) with transducer holder and
space for printer.
Warranty: The system and transducers should be covered with comprehensive onsite
warranty for 5 years.
Essential: 95% uptime guarantee should be provided
Manufacturer shall provide a loan system in case of machine failure.
List of installations within India in Govt. hospitals must be provided.