Specifications
F. No. 132/Anaes/RPC/SSK/12-13
Defibrilator with CPR Feedback – 01 No.
1. The defibrillator should be latest, lightweight, small size with bright colored display
2. The defibrillator should be with Rectilinear Biphasic technology, with 3 wave- form display on screen with
minimum 6.5 inches screen size.
3. Weight of defibrillator should not be more than 6.5 Kgs
4. It should display of both selected and delivered energy
5. It should have ability to energy selection from Paddles as well as unit.
6. In manual mode the unit should provide energy selection at (1-10, 15, 20, 30, 50, 70, 85, 100, 150, 200)
joules
7. It should have ability to measure chest compression rate and depth in real time with both visual & audible
feedback on screen should be available with CPR Index, Lapse time.
8. It should have both adult and pediatric paddles.
9. The unit should have transcutaneous external pacing with 40 milli-second pulse width
10. The unit should have facility for external as well as internal defibrillation.
11. The unit should do self test daily with facility to give print out of defibrillator testing report and also have
code ready indicator on unit.
12. It should have ability to filter out CPR artifacts and allowing person to see organized filtered rhythms
without interrupting chest compressions.
13. Battery should be capable of delivering minimum 3.5 hrs of ECG monitoring and capability of delivering
100 shocks once fully charged and should have charge level indicator on it.
14. The defibrillator should have facility for measuring following parameters
a. SPO2
b. ETCO2
c. NIBP
d. Temperature
15. Defibrillator should be able to upgrade for invasive monitoring.
16. Adult as well as pediatric saturation probes, and obese, adult and pediatric blood pressure cuff should be
provided
17. Facility for printed ECG should be there.
18. The defibrillator should be US FDA approved