Specifications
F. No. 131/Anaes/RPC/SSK/12-13
Radiant warmer – 02 Nos.
1. Floor Model Radiant Warmer with Ceramic heating element(s) with high-grade steel safety guard.
2. Non-visible infra-red radiation, therefore no impairment of visual examination.
3. Functions: it should have self test.
4. It should have intensity selection from 2 to 30 mW/cm
2
(depending on distance to the patient pad) in
increments of 2 mW/cm2.
5. Intensity selection should display in mW/cm2.
6. Visual and audible alarm in case of system faults and power failure.
7. Visual and audible alarm as well as automatic intensity reduction after 15 minutes above 10mW/cm2. User
should be able to confirm the high intensity for further 15 minutes. (Function can be switched-off
temporarily.)
8. Height-should be adjustable
9. It should have mobile stand with 4 anti-static castors, 2 of them with kickstop,
10. Mobile stand should be able to adjust according to the OT table width.
11. It should be adjustable to patient level 600 to 1000 mm, distance mark securing the safety distance to the
patient bed,
12. The rotation of warmer should be360°.
13. It should have vertical adjustment of ± 45°