F. No. 124/Anaes/RPC/SSK/12-13
Flash Autoclave – 02 Nos.
1. It should be a fully automatic table top high pressure steam sterilizer with chamber volume of 20 L for sterilization of
wrapped, unwrapped, hollow and porous materials and should have cycles for prions also.
2. It should have Integrated steam generator and vacuum pump.
3. It should be a B class & N class sterilizer with cycle time for wrapped instrument should be less than 25 minutes.
4. The sterilization chamber, jacket and steam generator should be made of high quality stainless steel
5. It should have an integrated water purification system.
6. It should have fully microprocessor control system for fully independent checking and control of the sterilization system.
Should be complete auto diagnostic system providing error codes on display in case of failure.
7. It should have thermal fuse protection against overheating of chamber and jacket and against operation of the steam generator
without water.
8. It should have LCD display and RS 232 interface. Should have automatic door locking mechanism.
9. Sterilization to be achieved at 134 degrees and 121 degrees . The sterilizer should have preloaded programs for wrapped
instruments, unwrapped instruments, textiles, rubbers and prions. It should also have preloaded test programs for Bowie-Dick
test and Vacuum test.
10. The sterilizer should come with standard accessories like sterilization trays, holders, start up kit etc.
11. It should be ISO, CE, TUV certified and should comply with EN 13060 standard for small sterilizers.
12. The company should have installation in india.