Specifications

F. No. 120/Pharma/RPC/SSK/12-13
2D Gel Electrophoresis with imaging System – 01 No.
Specifications
Complete system to separate, quantitate, characterize, visualize and analyze proteins in biological samples. System
should have following features:
A) Isoelectric focusing unit (1
st
Dimension):
System should include 12 different Individual Lane Control for running different samples, pH Gradients and focusing
protocols in a single run with different protocols.
Software Interface for creating and editing protocols and setting up the program rapidly.
Provision to interface with website for online data interpretation.
Must provide flexibility to run IPG strips gel side up, gel side down and with cup loading configuration.
System should have voltage 0–10,000 V, 1 V increments(50-10000V) or better
Current range should be 0–100 µA per lane at 1 µA intervals or better
System should have peltier based cooling platform.
Temperature range should be 10–25ºC ±1.0ºC @ max ambient 23ºC and 18–25ºC ±1.0ºC @ max ambient 31ºC.
Focusing trays should be made of polycarbonate for contamination free process.
System should be capable of simultaneously running minimum 12 IPG strips of various lengths like 7, 11, 13, 17, 18, and
24 cm.
System should have ramping Step, linear, gradual, and hold voltage ramping for each focusing step. Hold mode as a final
step to prevent diffusion when IEF is complete.
State-of the art Computer (if required to run the software) along with laser printer with UPS must be quoted along with the
system
B) SDS-PAGE Electrophoresis unit (2
nd
Dimension):
For Mini gels
(7 cm) capable of running 2 mini gels simultaneously
Capable of running hand cast as well as precast gel.
Separate running and casting module
Capable of using blotting module to do western blotting.
Leak proof, tape free and easy to assemble
Wet transfer blotting module of gel size should be provided
For Large Gels
Versatile system format to perform SDS, 1D, 2-D and Native gels
SDS-PAGE apparatus to run 11-13 cm gels and wet blotting system with features mentioned below.
Capable of running IPG strips
Provision to enable uniform pressure along entire length of the glass plates which is expected to provide leak proof
seal without tapes, grease or agarose.
Choice of glass plates, spacers & sandwich clamps to cast 2 gels.
Blotting module of similar gel size should be provided
Power supply
Output range (programmable) – 10-500 V, fully adjustable in 1 V steps and 0.01-2.5 A, fully adjustable in 1 mA
steps, 1-400 W power.
Type of output – Constant voltage, constant current, constant power or constant temperature.
Output terminals – 4 pair recessed banana jacks in parallel
Safety features - No-load detection, sudden load change detection, overload/short-circuit detection, over voltage
protection
Timer - 1 min–99 hr 59 min, fully adjustable