Co2 Incubator – 01 No.
Interiors must be made up of stainless steel or copper
The internal volume of the incubator should range between 200 – 250 L.
It must have minimum 6 door gas tight segmented access.
It should have dry heat sterilization option for decontamination.
Temperature control range between Ambient +3 to 55 Deg C or better
Recovery time at 37 Deg C must be 10 min or better
Humidity control 95% H or better
Recovery time for humidity 20 min or better
Carbon dioxide control range 0-20% at the inlet pressure of 12-15 LBS with the accuracy of 0.1 or better
Recovery time up to 5% in 5 min or better