F. No. 113/Pharma/RPC/SSK/12-13
Co2 Incubator – 01 No.
• Interiors must be made up of stainless steel or copper
• The internal volume of the incubator should range between 200 – 250 L.
• It must have minimum 6 door gas tight segmented access.
• It should have dry heat sterilization option for decontamination.
• Temperature control range between Ambient +3 to 55 Deg C or better
• Recovery time at 37 Deg C must be 10 min or better
• Humidity control 95% H or better
• Recovery time for humidity 20 min or better
• Carbon dioxide control range 0-20% at the inlet pressure of 12-15 LBS with the accuracy of 0.1 or better
• Recovery time up to 5% in 5 min or better