Specifications
F. No. 102/Bio/RPC/SSK/12-13
Deep Freezer -90°C – 01 No.
1. System should have Dual Refrigeration System and two independent compressors both units are controlled
independently and the two separate cooling system
2. Temp. Range : -90°C
3. Capacity : 820 L
4. Door type: Double Door
5. System should have Insulated inner doors
6. System should have easy to use door lock, should have heated vacuum valve to prevent a vacuum when
opening the door at short intervals
7. System should have Castors and door key lock
8. System should have Heated door frame for easy opening of door
9. System should have Micro processor controller with digital display
10. System should have 100% HCFC/CFC free
11. System should have Battery backup for alarm, logging and temperature display in case of power cut
12. System should have visual and acoustic alarm & Adjustable high/low temperature alarm
13. System should have Power failure alarm & Probe failure alarm
14. System should have Instrument failure alarm & Door open alarm
15. System should have Contact for external remote alarm & option for a for GSM alarm
16. System should have Integrated data logger with software included
17. System should have RS485/232 interface & USB read out
18. System should have auto cycle if probe failure & Ambient temp. Display Straws all alarms
19. System should have Direct download/upload on/from USB memory stick
20. System should have 3 level password protection
21. System should have Battery level indication & integrated memory
23. System should have Option for a chart recorder
24. System should have Power supply (V): 230 & Frequency (Hz): 50/60