F. No. 101/Bio/RPC/SSK/12-13
Deep Freezer -20°C – 01 No.
1. Temp. Range : -20°C
2. Capacity : 820 L
3. Door type: Double Door
4. System should have Insulated inner doors
5. System should have easy to use door lock, should have heated vacuum valve to prevent a vacuum when
opening the door at short intervals
6. System should have Castors and door key lock
7. System should have Heated door frame for easy opening of door
8. System should have Micro processor controller with digital display
9. System should have 100% HCFC/CFC free
10. System should have Battery backup for alarm, logging and temperature display in case of power cut
11. System should have visual and acoustic alarm & Adjustable high/low temperature alarm
12. System should have Power failure alarm & Probe failure alarm
13. System should have Instrument failure alarm & Door open alarm
14. System should have Contact for external remote alarm & option for a for GSM alarm
15. System should have Integrated data logger with software included
16. System should have RS485/232 interface & USB read out
17. System should have auto cycle if probe failure & Ambient temp. Display Straws all alarms
18. System should have Direct download/upload on/from USB memory stick
19. System should have 3 level password protection
20. System should have Battery level indication & integrated memory
21. System should have Option for a chart recorder
22. System should have Power supply (V): 230 & Frequency (Hz): 50/60