F. No. 96/Path/RPC/SSK/12-13
Flow Cytometer – 01 No.
- Air Cooled Argon ion Laser (488nm) and Red Diode Laser (635nm)
- The working principle should be single platform, fluorescence based flow cytometry.
- Capable of minimum 5 colour analysis.
- Have user – interchangeable optical filters
- Should have high performance forward scatter, side scatter and fluorescent detectors.
- Should be compatible to all routinely used flurochromes and upgradable in future.
- Should have auto sample tube loader for 30 or more tubes preferably with individual vortexing of tubes
- Should allow used defined statistics and gating with real time data analysis.
- Perform the following assays:- Absolute Total Cell Count, DNA Analysis, Multicolour Gating at least 4 colour
Immunophenotyping, Leukemia/Lymphoma, HLA B27
- System should be quoted with magnetic based cell sorter as an add on module for the flow Cytometry
- System should be upgradable to plate loader (96 Well plate ) in future
- The Company should have their own complete range of reagents for Leukemia (Acute/ Chronic ) as per WHO and
latest classification . Antibodies for the same has to be quoted.
- The company should provide hands on training for working on the flowcytometer (Application & Service support
from direct manufacturer)
- Start up kit should include sheath fluid, calibration reagent, reagent tubes, compensation kit, lyse solution.
- UPS & Colour Laser Printer to be quoted.