System information
__________________________________________________
Blastware Compliance Module
Blastware Operator Manual 4–53
b. If you would like to redefine the Setup File, press Cancel to exit the
Duplicate Setup dialog and return to the Monitor Scheduler. You
will have to return to the Compliance Setup dialog to make any
changes to the Setup File before returning to the Monitor Scheduler
feature.
c. If you select User Setup from PC or User Setup from Unit, or if the
file that you are transferring has a unique name, then the following
Save Series IV Monitor Scheduler dialog will appear:
Use descriptive filenames to help you identify these sessions for later
use (e.g. VDVSchedule_Day for daytime monitoring in the VDV
Record Mode).
The file extension, (*.SCH) is automatically added to the files as they
are saved. The default file location for saved .SCH is the
/Blastware/System/Dbase directory.
d. At this time, you will be prompted with the following dialog:
Select Yes if you want the unit to start using the Monitor Scheduler
settings the next time a recording session is initiated, either via
Blastware or as a manual key press on the unit. Select No if you don’t
want Monitor Scheduler to be enabled.
e. You are able to Enable or Disable the Monitor Scheduler feature on a
connected unit, at any time, by using the Unit > Monitor dialog. The
unit must be idle to access the Turn On or Turn Off button. In the
example below, the initial Current Monitor Mode: “Unit is idle.” and
it is not using a Monitor Scheduler session. The ability to Enable or
Disable the Monitor Scheduler is available by selecting the Turn On
or
Turn Off button: