Blastware Compliance Module
Blastware Operator Manual 4–31
d. Edit Compliance Setup ‘More’ for Micromate
Note: Edit Compliance Setup ‘More’ for Series II and III is covered in previous step.
(1) The More dialog within the Micromate Recording Setup dialog allows you to configure the
optional Auxiliary I/O for an optional external trigger or the optional Remote Alarm (Part No.
721A0901). Refer to the associated product specification sheet (Part No. 714U0901) that was
shipped with the Remote Alarm for more information. The Remote Alarm must be attached to the
optional Auxiliary Port (Part No. 721A0101) on the Micromate.
i Geophone Alarm Level
The alarm level controls the triggering of an optional Remote Alarm Controller, Part No.
721A0901 for Micromate.
The remote alarm can sound two alarms: Warning and Alarm. The Warning Alarm sounds when
Event activity exceeds the Geophone Alarm Level. The alarm level can be set to a different level
that is higher than the trigger level of the Geophone. This allows for the triggering of a second
alarm at the higher level.
For example, in pile driving operations the remote alarm triggers a warning when activity exceeds
the trigger level for the Geophone, notifying the operator that the pile driving activity has
exceeded that set level. The second alarm triggers when activity exceeds the alarm level, set to a
higher level, notifying the operator to re-evaluate the piling operation.
To set an alarm level, double click on the Geophone Alarm Level box. Enter the new level that is
within the stated range (Range: 12.570 – 254.000 mm/s).
ii Microphone Alarm Level
For a description of the optional remote alarm, see the Geophone channel in the previous section.
To set an alarm level, click on the Microphone Alarm Level box. Enter the new level that is
within the stated range (Range: 130 – 148 dB(L).
iii Auxiliary I/O
This command turns on the optional Auxiliary I/O feature, factory installed at the time of
purchase, and allows the monitor to be triggered by an External Trigger or a Remote Alarm. If
you do not want to use this feature, set Auxiliary I/O to Off. To enable the Auxiliary I/O, click
on the radio button next to the accessory that you will be using, External Trigger, such as a wire
make/break, or Remote Alarm.